Features and Benefits
Leading Hybrid Design
- The VS110 is a microscope-based system. Olympus has combined the optical leadership experience obtained through many decades and an innovative digital engine to create a virtual microscopy system second to none. With hundreds of Olympus scanning systems installed world- wide, the VS110 is the culmination of many years of engineering refinement.
The Whole Picture
- The careful integration of all the components creates a highly flexible system
- The slides can be scanned at true 20x/0.75 N.A. and 40x/0.95 N.A. with automated tissue detection and multiple regions of interest scanning possible. An option for 60x/1.35 N.A. and 100x/1.40 N.A. oil is available
- Virtual-Z, a multiple z-planes scanning and virtual focusing feature makes it ideal to use for scanning thick samples where depth of focus is important
High Capacity and Automatic Throughput
- The Olympus VS110 system can be fitted with a slide loader, with integrated barcode reader and hands-off scanning. This ability makes it quick and easy to archive slides, producing permanent records.
Friendly Operation
- With the Olympus VS110 users will be guided through the entire virtual slide acquisition process step-by-step by an intuitive Scan wizard. Full slide scanning can be set up with just 3 mouse clicks.
- With VS110 it is also possible to review the overview slide-images and the ability assign various regions of interest individual magnifications.
- The advanced software allows experienced users to be in full control of the details of the scanning process. Therefore, different scan settings can be assigned to individual slides, saving significant scanning time, as well as offering the possibility to only scan areas of interest, reducing the data file size.
Global Access
- VS110 images are stored and transmitted, a researcher can invite the rest of their group to view slides of interest, either individually or as a group, whether they are in the same building or situated around the world.
- The virtual facility offers controllable access for the user, such that with the correct permissions, they can add and edit files.
- The Conference mode allows a moderator to set up a conference so that all invited clients can see and discuss what the speaker is showing, plus it also records every aspect of the conference, such as the participants: names, notes and annotations.
Specifications:
|
|
VS110-L100 |
VS110-S5 |
VS110-S1 |
| Intended Specimen |
Observable
Specimen |
Glass slide with cover glass |
| Size of Glass Slide |
Width: 26mm(52mm in option), Length:76mm, Thickness:0.9mm-1.2mm |
| Size of Cover Glass |
Thickness: 0.12mm-0.17mm |
| Microscope Frame |
Illuminator |
Built-in Koehler illumination for transmitted light |
Objective Lenses
|
2X, 10X, 20X, 40X with a motorized revolving nosepiece |
| Optional 60X or 100X oil objectives |
| Motorized Stage |
Motorized XY stage with automatic control |
| Focusing |
Motorized automatic control |
| Digital Camera |
CCD Camera |
2/3� CCD camera, 6.45 μm x 6.45 μm pixel size,
high sensitivety, high resolution, Peltier cooled, |
| Image Correction |
Shading correction, Auto white balance |
| Slide Loading |
Manual Loading |
1 slide |
5 slides |
1 slide |
| Slide Loader |
100 slides |
None |
None |
| Scan |
Scan Area |
W 26mm x H 64 mm (W 52mm x H 64mm as an option) |
| Resolution |
Better than 0.33 micron/pixel when using 20X objective
Better than 0.16 micron/pixel when using 40X objective |
| System Control |
Computer |
Compatible with Windows XP Professional English Version |
| Network Interface |
100/1000 Mbit/s Ethernet |
| Memory |
4 GB RAM |
| Hard Disc Drive |
500 GB or more |
| Display |
High resolution 20.1� TFT monitor |
| Software |
Image format: vsi, JPEG, TIF
Zooming while scanning
Annotations
automatic sample detection
Z stack
Extended focus imaging
screen capture
Stepless zooming
Synchronised multi-images display
Automatic stitching
Slide loader control
Network Software with database function and conference up to 5 user accounts
consultation
Conference up to 20 user accounts, Conference more than 20 user accounts.
|
| Environment |
Power |
AC100-120V/220-240V, 50/60Hz |
| Weight *1 |
110Kg |
55Kg |
55 Kg |
Operating
Environment
|
Temperature: 15-28 degree centigrade |
| Power Consumption |
Less than 900VA |
|